AUDICOR 200 D Inovise Part Number 30047 (console part number 20158).
FDA Recall Z-0565-06 · Posted 2006-03-03
Advertisement
Recalling FirmInovise Medical, Inc.
Recall StatusTerminated
Product CodeDPS
Date Initiated2003-01-12
Date Terminated2006-07-24
Product Quantity28 units
Reason for Recall
Use of the AUDICOR 200 D in-line with a defibrillator/monitor resulted in interruption of pacing therapy to a patient when the AUDICOR 200 D stopped working due to low battery power.
Firm Action
On January 12, 2006, 2 of 3 customers were notified by telephone, with follow-up email. All three customers were visited by the firm on January 16-17, 2006. Customers were instucted to not use the device in-line with a defibrillator/monitor.
Root Cause
Other
Distribution Pattern
Product distributed to three EMS/Fire Departments located in Florida and Washington