Terminated

AUDICOR 200 D Inovise Part Number 30047 (console part number 20158).

FDA Recall Z-0565-06 · Posted 2006-03-03

Recalling FirmInovise Medical, Inc.
Recall StatusTerminated
Product CodeDPS
Date Initiated2003-01-12
Date Terminated2006-07-24
Product Quantity28 units

Reason for Recall

Use of the AUDICOR 200 D in-line with a defibrillator/monitor resulted in interruption of pacing therapy to a patient when the AUDICOR 200 D stopped working due to low battery power.

Firm Action

On January 12, 2006, 2 of 3 customers were notified by telephone, with follow-up email. All three customers were visited by the firm on January 16-17, 2006. Customers were instucted to not use the device in-line with a defibrillator/monitor.

Root Cause

Other

Distribution Pattern

Product distributed to three EMS/Fire Departments located in Florida and Washington