Terminated

RenovoCath RC 120; Temporary Intravascular Occluding Catheter Model number: MMRC120-DB-1111; Rx only. Cardiovascular: The RenovoCath RC120 Catheter is intended for the isolation of blood flow and delivery of fluids, including diagnostic and/or therapeutic agents, to selected sites in the peripheral vascular system.

FDA Recall Z-0565-2016 · Posted 2015-12-31

Recalling FirmRenovoRx, Inc.
Recall StatusTerminated
Product CodeMJN
Date Initiated2015-11-25
Date Terminated2016-02-29
Product Quantity11 units

Reason for Recall

Incorrect version of Instructions for Use were shipped to two US customers.

Firm Action

RenovoRx notified customers by phone and by mail of this issue. This notification requests that the customers remove the affected product from their inventory and remove and replace the incorrect Instructions for Use with the correct Instructions for Use provided by RenovoRx. The customers are requested to return the incorrect Instruction for Use to RenovoRx for disposition and are requested to complete a Recall Confirmation Form enclosed to document the completion of the requested Correction. For further questions, please call (650) 284-4433.

Root Cause

Labeling Change Control

Distribution Pattern

US Distribution to the states of : CA and FL.