CardioQuip Modular Cooler-Heater, Model: MCH-1000(i), Power: 115V-60Hz, 16A
FDA Recall Z-0565-2024 · Posted 2023-12-15
Advertisement
Recalling FirmCardioQuip, LLC
Recall StatusOpen, Classified
Product CodeDWC
Date Initiated2023-10-24
Product Quantity4 units
Reason for Recall
There is the potential that cooler-heater devices may contain bacterial contamination which could result in patient infection.
Firm Action
On October 24, 2023, CardioQuip issued a "Urgent: Medical Device Correction/Removal" notification via USPS to all affected consignees. CardioQuip ask consignees to take the following actions:
1. CardioQuip recommends that you remove your potentially contaminated MCH device from service immediately.
2. CardioQuip Customer Service will contact you by November 3, 2023 to provide your facility with a loaner device and to schedule an IWPR procedure for your MCH device.
3. Please complete and return the Receipt of Communication/Response Form attached to this letter to CardioQuip as soon as possible.
4. Any adverse events experienced relating to CardioQuip devices should be reported to CardioQuip or through the FDA s MedWatch Program:
Root Cause
Process control
Distribution Pattern
US: California, Texas, Pennsylvania, Washington, Ohio, Nebraska OUS: None
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.