Terminated

Accu-Chek Advantage blood glucose test strips; Reference numbers 03144917001, 03144704001, 03145263001, (sold within the United States) and 12030551001, 12030578001 and 12030586001 (sold outside of the United States).

FDA Recall Z-0567-06 · Posted 2006-02-28

Recall StatusTerminated
Product CodeLFR
Date Initiated2006-02-10
Date Terminated2007-02-08

Reason for Recall

There may be a thin piece of plastic across the mouth of the vial, preventing the cap from sitting snugly onto the vial and, if present, this situation will cause erroneously low or high blood glucose readings.

Firm Action

Notifications (letters) dated 2/10/06 were sent to healthcare professionals, retail pharmacies, hospitals and mail order firms asking them to provide their customers with information instructing the users to inspect their strip vials and to not use strips from defective vials. Mail order accounts were instructed to forward the recall notification to end users.

Root Cause

Other

Distribution Pattern

Nationwide, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Ecuador, England, France, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, New Zealand, Pakistan, Panama, Peru, Philippines, Poland, Russia, Singapore, South Africa, South Korea, Spain, Sweden, Taiwan and Thailand.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.