Datascope CS 100 Intra-Aortic Balloon Pump and Datascope CS 300 Intra-Aortic Balloon Pump, Datascope Corp. Mahwah, NJ 07430

FDA Recall Z-0567-2008 · Posted 2008-01-31

Recalling FirmDatascope Corp
Recall StatusTerminated
Product CodeDSP
Date Initiated2007-11-19
Date Terminated2008-05-20
Product Quantity70 units US

Reason for Recall

Defective Circuit Board -- A defect in the Printed Circuit Board's (PCB) in the CS 100 and CS300 Intra-Aortic Balloon Pump may cause the pump to exhibit intermittent malfunctions, which include failure to start-up or reset of the display screen during therapy.

Firm Action

Datascope sent a Product Field Correction notification letters to Hospital Administrators, via certified mail, on November 19, 2007. The letter informed them of a defect in the circuit board. It described the possible malfunction and provided instructions should malfunctioning occur. Datascope will be replacing the defective circuit boards in all affected IABP units. A Service Representative will be contacting the hospital to arrange for the replacement of the defective board.

Root Cause

Other

Distribution Pattern

Worldwide Distribution

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →