Terminated

Veta Peritoneal Catheter Kit, Curled, Ref No. M7225 Product Usage: Veta Peritoneal Dialysis Catheters are intended for acute and chronic access to the peritoneal cavity.

FDA Recall Z-0567-2015 · Posted 2014-12-12

Recalling FirmPfm Medical Inc
Recall StatusTerminated
Product CodeGCB
Date Initiated2014-10-20
Date Terminated2015-04-23
Product Quantity85 units

Reason for Recall

PFM Medical is recalling catheters and other medical devices because they may exceed USP limits for bacterial endotoxin.

Firm Action

PFm Medical sent an Urgent Device Recall Notification letter dated November 3, 2014 to all their customers who purchased the catheters and other medical devices. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete the recall acknowledgement/reply form and fax it to (760) 758-1167 or email it to [email protected]. Customers with any questions are instructed to contact Jessica Greene at (760) 758-8749.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution - US Nationwide in the states of MA, NC, UT, PA, TX, OH, LA and worldwide to: Germany and Malaysia.