Meridian Bioscience issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees beginning on 10/15/2025 via email. The notice explained the problem with the device, health hazard, and requested the following:
"Required Actions:
" Identify remaining inventory of Alethia C. difficile, Lot 480050U036.
" Identify the Reaction Buffer bag located within the kit box.
" Remove Reaction Buffer Tubes from Bag and inspect each cap for the presence of a gasket.
" Segregate Reaction Buffer Tubes which have no gasket present. Record the quantity of tubes with no gasket present on the CONFIRMATION OF NOTIFICATION Form.
" Discard Reaction Buffer Tubes which have no gasket present in accordance with local, state, or federal regulations.
Note: Execution of the test procedure requires the use of a Reaction Buffer tube. The remaining components in this kit may continue to be utilized for patient testing so long as a new Reaction Buffer tube is available. Once the Reaction Buffer is depleted, the remaining kit components must be discarded in accordance with local regulations.
" Meridian Bioscience Inc. will provide credit for the number of tests impacted.
" Complete and return the CONFIRMATION OF NOTIFICATION Form to: Product Support Manager, Meridian Bioscience, Inc., 3471 River Hills Drive, Cincinnati, OH 45244; Fax: (513) 272-5432 or Email:
[email protected]. Complete this form if there is no remaining inventory."
For any question regarding this notification please call Meridian Bioscience Technical Services at 1-800-343-3858, or email at
[email protected].