Terminated

Coulter LH 700 Series and GEN S Hematology Analyzers 864.5220 Automated differential cell counter

FDA Recall Z-0571-03 · Posted 2003-02-25

Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Date Initiated2002-12-10
Date Terminated2004-04-21
Product Quantity447

Reason for Recall

Device design allows for cross contamination issue to occur, possibly affecting assays and results reports.

Firm Action

Firm will check and shorten sweepflow tubing on a next visit basis during routine service visits.

Root Cause

Other

Distribution Pattern

Nationwide and Canada