Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004
FDA Recall Z-0571-2014 · Posted 2013-12-24
Advertisement
Recalling FirmData Innovations, Inc.
Recall StatusTerminated
Product CodeJQP
Date Initiated2013-11-06
Date Terminated2014-12-05
Product Quantity24 systems
Reason for Recall
When connected to the Sysmex UF-100, the
Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003,
and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error
Firm Action
Data Innovation notified consignees by Important Medical Device Recall Notification dated 11/5/13 to all list serv recipients of the problem and posted critical alert on Customer Web Portal. A certified letter was issued to impacted customers and consignees if no response was recieved from e-mail.
Root Cause
Software design
Distribution Pattern
Nationwide Canada and Britain