Exactech 12/14 Biolox Alumina Femoral Head 28MM +3.5, hip prosthesis component, Exactech, Gainesville, FL 32653
FDA Recall Z-0572-2008 · Posted 2008-03-12
Advertisement
Recalling FirmExactech, Inc.
Recall StatusTerminated
Product CodeLPH
Date Initiated2007-11-01
Date Terminated2012-05-06
Product Quantity10
Reason for Recall
Mislabeled: A ten piece lot of 140-32-03, 12/14 BIOLOXforte Alumina Femoral Heads 32mm (1069020 - 1069029), from Ceramtec were incorrectly labeled as 140-28-03 12/14 BIOLOXforte Alumina Femoral Heads, 28mm
Firm Action
Exactech issued a recall letter, titled Important Product Recall Notice on 10/31/2007 to the Distibutors that received the recalled product. It requested that all device recipients be notified and all product be returned to the firm.
Root Cause
Process control
Distribution Pattern
Distributed to NY, OH and TN.