Smith & Nephew GENESIS(R) II/LEGION(TM) PATELLAR COMPONENT, RESURFACING ROUND, 32MM STANDARD, FOR CEMENTED USE ONLY WITH JOURNEY PEG, STERILE EO, Rx only, REF Number: 71934490 knee implant.
FDA Recall Z-0573-2016 · Posted 2016-01-04
Advertisement
Recalling FirmSmith & Nephew, Inc.
Recall StatusTerminated
Product CodeJWH
Date Initiated2015-11-09
Date Terminated2016-03-17
Product Quantity20 units
Reason for Recall
One lot of 35 mm patella implants is labeled as 32 mm. These implants have pegs which are spaced proportionally; therefore the mismatch prevents the pegs from properly aligning with pre-drilled holes during implantation.
Firm Action
The firm, Smith & Nephew Inc, notified the distributors by Fed- Ex on 10/20/2015 and sent an "Urgent - Product Recall 1st Notification - Urgent R-2015-22" letter dated 11/9/2015 to their customers. The letter described the product, problem, and actions to be taken. The customers were instructed to inspect your inventory and locate any unused devices, quarantine them immediately, complete and return the Inventory Return Certification Form by fax to: +1-901-566-7975 or email to: [email protected], even if you do not have the product, contact Smith & Nephew's Field Actions Department via email or fax (above) to obtain a return authorization (RA) number, and return any affected product to Smith & Nephew, Attn: Global Field Actions, Global Distribution Center, 3303 E Holmes Road, Memphis, TN 38118.
If you have any questions, call Director, Regulatory Affairs 901-399-5520 or email: [email protected].
Root Cause
Process design
Distribution Pattern
US Distribution to states of:FL, MO, IN and TX.