BioTek brand Automated Microplate Reader; model ELx808, Roche catalog number 03145387001.
FDA Recall Z-0574-05 · Posted 2005-03-08
Advertisement
Recalling FirmRoche Diagnostics Corp.
Recall StatusTerminated
Product CodeJJQ
Date Initiated2005-02-15
Date Terminated2006-05-10
Reason for Recall
The parameters loaded in the analyzer for the hepatitis C virus do not match those in the package insert, and patient results may be reported out as false positives or false negatives.
Firm Action
Recall letters dated 2/15/05 are being sent to each consignee instructing them to discontinue use of pre-programmed HCV parameters with the Amplicor heatitis C virus test and that previous test results within a specific range be reviewed by a physician.
Root Cause
Other
Distribution Pattern
Nationwide and Canada.