Access Immunoasay Systems Digoxin Reagent
FDA Recall Z-0576-05 · Posted 2005-03-08
Advertisement
Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeKXT
Date Initiated2005-02-23
Date Terminated2005-09-02
Product Quantity4352 domestically.
Reason for Recall
Variability in results on Patient and QC results of 20-25%, high and low.
Firm Action
Letters were sent to customers via UPS overnite, they request that alternative methods be used for patient testing until new material is made available. Customers are requested to ask for replacements using a fax back form.
Root Cause
Other
Distribution Pattern
Nationwide and Canada