Roche COBAS 6000 Clinical Chemistry analyzer; Catalog Number: 04745914001, Roche Diagnostics Corporation, Indianapolis, IN 46250.

FDA Recall Z-0576-2008 · Posted 2008-03-04

Recall StatusTerminated
Product CodeJJY
Date Initiated2007-12-04
Date Terminated2008-08-04
Product Quantity1

Reason for Recall

Incorrect results: CRP assays may be calculated using incorrect calibration parameters which could result in falsely high or falsely low patient results being reported.

Firm Action

Roche Diagnostics contacted the consignee by phone on 12/4/07 and by follow-up letter dated 12/4/07. The user was instructed to discontinue use of two CRP assays on this analyzer. The letter described a workaround and that a new revision of software, due out first quarter 2008, will fix this.

Root Cause

Software design

Distribution Pattern

USA Distribution in the state of Nebraska.