Terminated

Appraise DBS Test Pak/HW-Kronos KROJI, Product K8603, each kit packed in a bag or envelope and contains 1 Triad Medium Alcohol Prep Pad, Cat. No. 10-3001, NDC 50730-3001-0, blood specimen collection supplies and instructions. Blood specimen collection

FDA Recall Z-0580-2012 · Posted 2012-01-11

Recall StatusTerminated
Product CodeLCP
Date Initiated2011-01-09
Date Terminated2012-12-26
Product Quantity152 kits

Reason for Recall

Kits contain recalled Triad alcohol pads

Firm Action

Heritage Labs sent a Device Notification and Safety Alert letter dated January 12, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Heritage Labs sent replacement alcohol prep pads from a different supplier not affected by the recall to user of Heritage Labs kits. Heritage Labs recommended the replacements be on "as needed" basis. Upon opening kit, users should immediately dispose of sealed TRIAD brand alcohol pads. Users are to use the replacement wipes instead of the wipes contained in the kits. For any questions call (888)764-4120.

Root Cause

Material/Component Contamination

Distribution Pattern

Worldwide Distribution - USA (nationwide) and Germany

Related FDA 510(k) Records