Rebel Monorail PtCr Coronary Stent System, REF / Catalog No. H7493927324300, Rebel MR OUS 3.0mm x 24mm. The Rebel Monorail (MR) Stent System is a balloon expandable Platinum Chromium (PtCr) Alloy stent pre-mounted on a Monorail Catheter.

FDA Recall Z-0580-2020 · Posted None

Recall StatusTerminated
Product CodeMAF
Date Initiated2019-10-24
Date Terminated2023-03-20
Product Quantity14,145 devices total

Reason for Recall

Certain dilation and extension catheters are being recalled because these catheters may have missed the hydrophilic coating step, which creates a lubricious surface on the devices, during the manufacturing process.

Firm Action

The firm, Boston Scientific, sent an "Urgent Medical Device Removal" letter dated 10/28/2019 overnight mail delivery to its customers. The letter stated the issue, identified affected product and stated that product should not be used. The customers were instructed, if any product is identified to segregate immediately and return to Boston Scientific. The customers were also instructed to Immediately complete the RECALL REMOVAL REPLY VERIFICATION TRACKING FORM and Scan/e-mail to: [email protected] OR Fax to #: 1-866-213-1806. This form must be completed and returned in all cases even if you do not have any affected product. If you are a distributor, please note that the removal notification should be forwarded to your customers. If you have any questions, contact Boston Scientific Quality Systems at 763-494-1133 or email: [email protected].

Root Cause

Process control

Distribution Pattern

Nationwide including Washington D.C. and Puerto Rico. OUS to include Canada.