UniCel Dxl 800 Access Immunoassay System Software Versions 2.2.1 and earlier
FDA Recall Z-0583-06 · Posted 2006-03-03
Advertisement
Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeJJE
Date Initiated2006-02-06
Date Terminated2012-03-06
Product Quantity280 total (263 in US & 17 in Canada)
Reason for Recall
The UniCel Dxl 800 Access Immunoassay may aspirate a sample from an incorrect rack and generate a result, which is not flagged (under a specific set of circumstances)
Firm Action
A Product Corrective Action letter will be mailed the week of Feb 06 2006 to all UniCel 800 Dxl customers informing them of a rare condition on the UniCel Dxl 800 Access Immunoassay Systems. Under a specific set of circumstances, the UniCel Dxl 800 may aspirate a sample from an incorrect rack and generate a result, which is not flagged. (Letter is sent by US mail)
Root Cause
Software design
Distribution Pattern
Nationwide and Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.