Terminated

In-the-Ear hearing aids, Model Numbers: T9ITC-DW-LP, FT6ITC-D-MP, FT6ITE-D-MP, FT8CIC-MP, FT8ITC-W-HP, FT8ITC-W-LP, FT8ITC-W-MP, FT8ITE-DW-HP, FT8MIH-W-HP, FT8MIH-W-MP, LT5CIC-MP, LT5CIC-UP, LT5IIC-LP, LT5ITC-DW-MP, LT5ITC-DW-UP, LT5ITE-DW-HP, LT5ITE-DW-MP, LT5ITE-DW-UP, LT5MIH-W-HP, LT5MIH-W-MP, LT5MIH-W-UP, LT7CIC-HP, LT7CIC-LP, LT7CIC-MP, LT7CIC-W-MP, LT7IIC-LP, LT7ITC-DW-LP, LT7ITC-DW-MP, LT7ITC-MP, LT7ITC-W-HP, LT7ITC-W-MP, LT7ITE-DW-HP, LT7ITE-DW-MP, LT7ITE-DW-UP, LT7ITE-W-UP, LT7MIH-W-HP, LT7MIH-W-LP, LT7MIH-W-MP, LT9CIC-HP, LT9CIC-MP, LT9CIC-W-MP, LT9IIC-LP, LT9ITC-DW-HP, LT9ITC-DW-LP, LT9ITC-DW-MP, LT9ITE-DW-HP, LT9ITE-DW-MP, LT9ITE-DW-UP, LT9MIH-HP, LT9MIH-MP, LT9MIH-W-HP, LT9MIH-W-LP, LT9MIH-W-MP, TST17IIC-LP, TST17ITC-DW-HP, TST17ITE-HP, TST17MIH-W-UP, TST6CIC-HP, TST6IIC-LP, TST6ITC-MP, TST6ITC-W-HP, TST6ITC-W-MP, TST6ITE-DW-UP, TST6ITE-HP, TST6ITE-W-UP, TST6MIH-W-HP, TST9ITC-DW-MP, TST9ITE-DW-HP, TST9ITE-MP, TST9MIH-W-MP, VA8CIC-HP, VA8CIC-MP, VA8IIC-LP, VA8ITC-DW-HP, VA8ITC-DW-MP, VA8ITC-W-MP, VA8ITE-D-HP, VA8ITE-DW-HP, VA8ITE-DW-UP, VA8MIH-W-MP

FDA Recall Z-0584-2020 · Posted None

Recalling FirmGN Hearing A/S
Recall StatusTerminated
Product CodeESD
Date Initiated2019-09-18
Date Terminated2023-05-31
Product Quantity285 units

Reason for Recall

GN Hearing received reports from our manufacturing site in US regarding a software bug in our service & repair tool which affected the capture and restoration of user data in a limited number of repaired hearing aids. The software bug caused these repaired hearing aids to contain altered specifications for sound amplification, potentially putting users residual hearing at risk of experiencing unintended additional available gain to varying degrees i.a.18dB flat linear gain or potentially in worst case Full On Gain. However, no injuries have been reported.

Firm Action

The firm initiated the recall by telephone on 09/18/2019 and followed with a letter customized to each consignee. The requests the return of the affected units. The letter also requests that any entity that distributed the product notify their customers.

Root Cause

Other

Distribution Pattern

Nationwide