The LeadCare II Blood Lead Test System relies on electrochemistry and a unique Sensor to detect lead in whole blood. The LeadCare II Test Kit is provided with the following materials: " Sensors (2 Containers of 24 sensors each; 48 tests) " Treatment Reagent Tubes (2 packages of 24 tubes each; 48 tests) " Heparinized Capillary Tubes/Plungers (50) " Transfer Droppers (50) " Calibration Button (1) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)
FDA Recall Z-0584-2026 · Posted 2025-11-21
Advertisement
Recalling FirmMagellan Diagnostics, Inc.
Recall StatusOpen, Classified
Product CodeDOF
Date Initiated2024-02-23
Product Quantity1,774 units
Reason for Recall
Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.
Firm Action
On March 3, 2024, MEDICAL DEVICE FIELD CORRECTION letters were sent to customers. Immediate Actions:
1. Identify existing inventory of impacted product and suspend its distribution. 2.
Identify customers that received the impacted product and notify them about this field action. 3. Return remaining impacted product to Magellan Diagnostics, Inc. for destruction. To initiate the return process, ensure that you note the quantity of product to be returned on your Confirmation of Notification Form. Actions to Be Taken by Magellan: 1. Magellan Diagnostics will continue to investigate the cause of the labeling error and initiate appropriate preventive actions. 2. Magellan will replace unexpired impacted product upon request. Product will be replaced based on availability, and on a first-come first-served basis.
Root Cause
Labeling mix-ups
Distribution Pattern
US Nationwide distribution.