Boston Scientific SlimLine 200 Micron Holmium Fiber Delivery Device, Catalog No. 840-840, Sterile, Made In Israel, Manufactured By LUMENIS, P.O.Box 240, Yokneam, ISRAEL.
FDA Recall Z-0585-2008 · Posted 2008-04-09
Advertisement
Recalling FirmLumenis, Inc.
Recall StatusTerminated
Product CodeGEX
Date Initiated2007-12-18
Date Terminated2008-04-09
Product Quantity59 units.
Reason for Recall
Incorrect Label: The Units outer pouch is labeled 200 micro fiber however, the inner pouch's label affixed to the product indicated that it is a 365 micron fiber. The micron fiber contained in the inner pouch is 200 micro fiber. The inner product is mislabeled with the wrong size.
Firm Action
The firm notified Consignee by certified letter on 12/18/2007, and asked that all affected products be return. Any questions concerning this recall can be addressed by contacting the Manager, Regulatory Compliance, at 1-877-586-3646 ext. 3355.
Root Cause
Labeling mix-ups
Distribution Pattern
MA
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.