4Fr x 20cm Single Lumen CT Midline Basic Tray, Product Code: VYML4S1001

FDA Recall Z-0585-2020 · Posted None

Recall StatusTerminated
Product CodePND
Date Initiated2019-09-24
Date Terminated2020-10-21
Product Quantity75

Reason for Recall

Potential that 3Fr size catheter placed in the catheter tray instead of 4Fr for the 4Fr x 20cm Single Lumen CT Midline Basic Tray.

Firm Action

On September 25, 2019 the firm distributed Urgent Medical Device Recall Communication letters by e-mail instructing them to: 1. Check your stock for the products included within the scope of the recall. Immediately cease use and distribution, and quarantine all affected product immediately. 2. Count and document your affected inventory on the Recall Acknowledgement and Inventory Return Form in Attachment 1. Please complete and return the document even if you currently have no product on hand. 3. Fax or email the form to Vygon including customer contact information. 4. Vygon customer service will contact the customer and issue a Return Merchandise Authorization number. 5. Return the product as well as the original signed Recall Acknowledgement and Inventory Return Form using the instructions from Customer Service involving product pick-up. 6. All inventory will be credited or replaced at no cost to you. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax. If you have any questions, please contact your local sales representative or Vygons Customer Service Department at 1-800-473-5414 Monday-Friday 8am 6pm, or by e-mail at [email protected].

Root Cause

Packaging process control

Distribution Pattern

One Customer Located in TN

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.