MultiDiagnost 4 - x-ray system, multi functional/universal system for general use.

FDA Recall Z-0586-03 · Posted 2003-02-27

Recall StatusTerminated
Product CodeJAA
Date Initiated2003-01-10
Date Terminated2005-04-11
Product Quantity22 units

Reason for Recall

INTERNAL SHORT CIRCUIT - MAY ALLOW FOR STAND AND TABLE TO MOVE SPONTANEOUSLY IN UPWARD OR DOWNWARD MOVEMENT

Firm Action

The firm issued a letter dated January 10, 2003 to their customers advising of the problem and that a respresentative would be contacting customers to arrange a time to visit to install new interface board. This is being conducted under FCO 70800007

Root Cause

Other

Distribution Pattern

The firm distributed to 22 hospitals/medical centers located in CA, FL, IA, IL, IN, KY, MN, MO, NC, ND, NY, OH, SC, TN, TX, VA, and WV.