Evolution Loading Car and Transfer Carriage- Accessory to the Sterilizer, Steam
FDA Recall Z-0586-2020 · Posted None
Advertisement
Recalling FirmSteris Corporation
Recall StatusTerminated
Product CodeFLE
Date Initiated2019-10-22
Date Terminated2020-06-09
Product Quantity237 units
Reason for Recall
The locking pin of the front wheel asse mbly of the
Evolution Transfer Carriages may not remain full y engaged with the transfer carriage and result in the
transfer carriage and loading car unexpectedly falling forward
Firm Action
Steris Corp. issued URGENT DEVICE FIELD CORRECTION dated 10/22/19 informing users reason for recall, and action to take: A STERIS Service Technician will contact your facility to repair your unit(s). Users can continue to use the transfer carriage prior to this repai r. Users are advised to use caution
when operating the transfer carriage, and avoid bumping into other equipment or rolling over uneven surfaces.
If you have questions regarding this matter, please contact
STERIS Technical Support, at 800-333-8828, STERIS Customer Service at 1-800-548-4873, or your local STERIS Representative.
Root Cause
Process control
Distribution Pattern
Nationwide Foreign: Canada, Peru, Mexico, Brazil, Saudi Arabia, Columbia