MOVES System Lithium Polymer Batteries The MOVES System is a portable computer controlled electrically powered transport emergency ventilator intended to provide continuous or intermittent ventilatory support for the care of adults who require ventilation.
FDA Recall Z-0587-2016 · Posted 2016-01-11
Recalling FirmThornhill Research Inc
Recall StatusTerminated
Product CodeBTL
Date Initiated2015-11-19
Date Terminated2019-04-12
Product Quantity178 units
Reason for Recall
Batteries found unresponsive after storage, not recognized by main device, and are perpetually enabled.
Firm Action
Thornhill Research sent a Recall Notification letter dated December 15, 2015 to affected customers. The letter identified the affected products, problem and actions to be taken. Customers were instructed to update their device software.
For questions customers were instructed to call 416-597-1325.
Root Cause
Component design/selection
Distribution Pattern
Worldwide Distribution: US distribution in state of: Virginia and to the country of: Singapore.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.