Dimension LIP Flex Reagent Cartridge, Catalog# DF55A
FDA Recall Z-0588-2011 · Posted 2010-12-10
Advertisement
Recalling FirmSiemens Healthcare Diagnostics, Inc.
Recall StatusTerminated
Product CodeCHI
Date Initiated2010-03-10
Date Terminated2011-08-04
Product Quantity21171
Reason for Recall
Open Well instability: After calibration, QC shifts high (out of range) after the Flex Well is hydrated. Shifts of up to 20% have been observed in both QC and patient samplesas soon as two hours post hydration.
Firm Action
Recalling firm issued "Urgent Field Safety Notice", dated March 2010, informing users of the problem and to transition to the revised Lipasae method.
Root Cause
Other
Distribution Pattern
Nationwide Distribution throughout the US.