MEVION S250, used for proton radiation therapy.
FDA Recall Z-0588-2015 · Posted 2014-12-12
Advertisement
Recalling FirmMevion Medical Systems, Inc.
Recall StatusTerminated
Product CodeLHN
Date Initiated2014-11-10
Date Terminated2015-03-18
Product Quantity1
Reason for Recall
Software defect that causes an incorrect dose compensation function to be applied to the internal dose ionization chamber when either pressure or temperature sensor malfunction. This could result in an incorrect dose delivery of no more than 5%.
Firm Action
Mevion Medical Systems sent an Important Safety Notice letter to the affected customer. The letter identified the affected product, problem and actions to be taken. A software update addressing the issue will be provided to the customer on or about December 5, 2014.
For questions or concerns, email [email protected]
or call 978-540-1555.
Root Cause
Software design
Distribution Pattern
US Distribution in the state of MO