GE LOGIQ E10 Ultrasound System, Model #530000E10. The device is intended for use by a qualified physician for ultrasound evaluation of Fetal/Obstetrics; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculoskeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (vascular).

FDA Recall Z-0588-2019 · Posted None

Recalling FirmGE Healthcare, LLC
Recall StatusTerminated
Product CodeIYN
Date Initiated2018-09-14
Date Terminated2019-05-29
Product Quantity103 devices(64 US; 39 OUS)

Reason for Recall

There is a potential for an undetected over temperature condition on the trans-esophageal (TEE) probe leading to a possible burn injury.

Firm Action

The recalling firm, GE Healthcare, issued an "URGENT MEDICAL DEVICE CORRECTION" letter dated 9/14/2018 via traceable means to each of their customer locations. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: Do not use a TEE probe (GE model 6Tc-RS) from another device on the LOGIQ E10 system until your LOGIQ E10 is corrected. You may scan normally with any other probes. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Root Cause

Process change control

Distribution Pattern

Worldwide Distribution: US (nationwide) to states of: AZ, IL, MA, ME, MN, MT, PA, TX, and WA and countries of: Foreign distribution was made to Austria, Denmark, Finland, and Norway.