Terminated

BD PrecisionGlide Needle 30G x 1/2 (0.3 mm X 13mm) REF 305106 - Product Usage: used for general purpose injection and aspiration of fluids from vials, ampoules and parts of the body below the surface of the skin.

FDA Recall Z-0588-2021 · Posted None

Recall StatusTerminated
Product CodeFMI
Date Initiated2020-11-18
Date Terminated2024-02-15
Product Quantity1,430,000

Reason for Recall

BD has confirmed the presence of blood on the cannulas of the BD PrecisionGlide needles 30G X 1/2" Lot No. 9128897. The possibility of infection may be present if the blood is not noticed prior to use by the clinician.

Firm Action

Customers are provided with the following actions to take in the firm's recall notification letter: 1. Immediately review your inventory for the specific Catalog (Ref) and lot number listed above. Destroy all product subject to the recall following your institutions process for destruction. 2. Share this recall notification with all users within or outside your facility network to which you may have further distributed the affected product to ensure they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement. 4. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fishers Lane, Rockville, MD 20852-9787 Replacements will be provided for all discarded inventory. If you require further assistance, please contact: BD Contact: Customer/Technical Support US Contact Information: (844) 823-5433 MonFri 8:30am and 5:00pm (EST) Areas of Support: For customers outside the US, contact your local BD representative or distributor

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of CA, CT, FL, IL, KY, MO, NJ, NY, OH, PA, PR, SC, TN, VA and the countries of Canada, China, Singapore.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.