Maquet Servo Ventilator SV900C/D/E (SV900) Siemens-Elema AB, Solna, Sweden
FDA Recall Z-0589-2010 · Posted 2010-01-05
Advertisement
Recalling FirmMAQUET Inc.
Recall StatusTerminated
Product CodeCBK
Date Initiated2009-11-17
Date Terminated2012-06-04
Product Quantity29364 (8438 of those in US)
Reason for Recall
Some Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the ventilator's pressure transducers.
Firm Action
Maquet sent Device Correction letters commencing 11/17/09 and ending 11/23/09 by Federal Express.
Root Cause
Device Design
Distribution Pattern
Worldwide distribution