Terminated

GE Responder 2000 Defibrillator/Monitor, models 2025653, 2026109, and 2026114 that were manufactured between July 25, 2008 and October 2, 2009 by GE Healthcare.

FDA Recall Z-0589-2011 · Posted 2010-12-10

Recall StatusTerminated
Product CodeMKJ
Date Initiated2010-04-28
Date Terminated2011-12-08
Product Quantity130 units

Reason for Recall

Some units may be non-operational during a rescue, due to a defective PCB interconnect on traces connected to the reed relay. This will result in an inability to analyze the patient's heart rhythm or deliver therapy during a rescue.

Firm Action

Cardiac Science issued a Urgent Medical Device Corrective Action letter dated April 28, 2010 requesting GE Healthcare cease distribution of affected devices and to instruct end users of distributed devices to schedule a service call to have the device corrected. Cardiac will send rework kits no later than June 30, 2010. CSC can be contacted at 44 161 926 0011 to arrange return material authorization and replacement of affected devices.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution: Throughout the US, and in the countries of Germany, Korea, Peru, Singapore, and Spain.