BD Vacutainer Push Button Blood collection Sets with Pre-Attached Holder; Catalog Number: 368656; 23 g X12 " PBBCS w/PAH (Preattached), BD Diagnostics, Franklin Lakes, NJ
FDA Recall Z-0590-2008 · Posted 2008-01-23
Advertisement
Recalling FirmBecton Dickinson & Company
Recall StatusTerminated
Product CodeJKA
Date Initiated2007-11-20
Date Terminated2008-09-08
Product Quantity769,200 USA; 1,300 Canada
Reason for Recall
Safety Mechanism Failure: the safety mechanism of the BD Vacutainer Push Button Collection Set with Pre-attached holder cannot be activated; thus the needle cannot retract into the rear barrel.
Firm Action
BD sent Recall notification letters, Urgent Class II Product Recall, to customers and distributors on 12/10/07 through a third party, West Inc., Appleton, Wi. Mailings were sent return receipt requested. Letters requested immediate inventory search and product return. Contact number for questions was given as 1-800-631-0174
Root Cause
Material/Component Contamination
Distribution Pattern
Worldwide-USA and Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.