Microtek Medical, Inc., Microscope Drape for Zeiss Standard Model/Catalog Number: 70890655 Product Description: The Microtek Medical, Inc., Microscope Drape for Zeiss Standard is intended is to protect surgical microscopes from contamination during various procedures. Labeling for the product is as follows: (1.) INNER CARTON LABEL: Distributed by; MICROTEX, MICROSCOPE DRAPE FOR ZEISS STANDARD, 20 x 66 (51cm x 168 cm), REF 70890655, QTY 12, LOT DA102032, Microtek Medical, Inc., Columbus, MS 39702, www.microtekmed.com, CE0044, EC, REP, Microtek Medical B.V. Zutphen, The Netherlands, Manufactured in the Dominican Republic of U.S. Materials, STERILE, EO, 2015-07, (01)30748426064206, (17) 150700(10) DA102032, INNER, Rev. C. (2.) PRODUCT LABEL: Distributed by; MICROTEX, MICROSCOPE DRAPE FOR ZEISS STANDARD, 20 x 66 (51cm x 168 cm), REF 70890655, QTY 1, LOT DA102032, Microtek Medical, Inc., Columbus, MS 39702, www.microtekmed.com, CE0044, EC, REP, Microtek Medical B.V. Zutphen, The Netherlands, Manufactured in the Dominican Republic of U.S. Materials, STERILE, EO, 2015-07, (01) 00748426064205, (17) 150700(10) DA102032, EACH, Rev. D. (3.) CARTON LABEL: Distributed by; MICROTEX, MICROSCOPE DRAPE FOR ZEISS STANDARD, 20 x 66 (51cm x 168 cm), REF 70890655, QTY 12/CS, LOT DA102032, Microtek Medical, Inc., Columbus, MS 39702, www.microtekmed.com, CE0044, EC, REP, Microtek Medical B.V. Zutphen, The Netherlands, Manufactured in the Dominican Republic of U.S. Materials, STERILE, EO, 2015-07, (01)50748426084200, (17) 150700(10) DA102032, INNER, Rev. D.

FDA Recall Z-0592-2011 · Posted 2010-12-10

Recall StatusTerminated
Product CodeKKX
Date Initiated2010-08-23
Date Terminated2010-12-14
Product Quantity5 Cases, 12 drapes per case, 60 Drapes total distrubuted

Reason for Recall

This product is being recalled due the product is miss-labeled as being sterilized by EO processing; however, the product was processed using Gamma sterilization.

Firm Action

On 08/23/10 Microtek Medical, Inc., Jacksonville, Florida notified all of their customers via a telephone call and requested immediate return of the product . The firm followed up with each of the customers by sending out an Urgent: Medical Device Recall letter (written notice), dated September 02,2010, to each on 9/8/10. The letter identified the affected product, stated the reason for the recall, and asked that the product be returned immediately. Customers are to complete and return the enclosed response form. Anyone needing further assistance is asked to contact Elizabeth Jolly, V.P. Legal at 1-888-844-0988.

Root Cause

Error in labeling

Distribution Pattern

Nationwide Distribution