Terminated

Arsenal Spinal Fixation System, Set Screw, Part No. 47127 Product Usage: Usage: The Arsenal Spinal Fixation System is intended for posterior, non-cervical, spinal fixation as an adjunct to fusion for the treatment of degenerative disease, deformity, and trauma indications. The Arsenal System consists of a variety of shapes and sizes of rods, screws, hooks, connectors, and bridges that provide temporary internal fixation and stabilization during bone graft healing and/or fusion mass development. The screws, hooks, connectors, and bridges are manufactured from surgical grade titanium alloy (Ti-6AI-4V ELI). The rods are available in commercially pure titanium, titanium alloy, and cobalt chrome (CP Ti Grade 4, Ti-6Al-4V ELI, and Co-28Cr-6Mo).

FDA Recall Z-0592-2017 · Posted 2016-11-16

Recalling FirmAlphatec Spine, Inc.
Recall StatusTerminated
Product CodeNKB
Date Initiated2016-10-05
Date Terminated2017-02-17
Product Quantity16,972 units

Reason for Recall

Alphatec Spine is recalling the Arsenal Spinal Fixation System Set Screw due to a trend in set screw postoperative disengagement from the screw body.

Firm Action

Alphatae Spine sent an Urgent Medical Device Recall notification letter dated September 29, 2016 to customers. The letter identified the affected product, problem and the actions to be taken. Customers are instructed to fill out and return the last page of recall letter using one of the methods identified in the letter.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide Distribution - US Nationwide in the countries of Japan, UK, NSW, Spain.