Vertical drive brake (on patient supports). Used on certain Philips Medical Systems MX series and Brilliance series CT scanners. The device is a vertical drive brake that controls the vertical position of a patient support system or couch. The patient support system is intended to be used to place the patient in the correct horizontal and vertical position in order to be scanned. The couch is an accessory product to a computed tomography scanner.

FDA Recall Z-0597-2009 · Posted 2009-01-16

Recall StatusTerminated
Product CodeKPS
Date Initiated2008-09-23
Date Terminated2012-03-01
Product Quantity559

Reason for Recall

The product is being recalled because the patient support may travel downwards without being commanded to move.

Firm Action

A On 11/15/2007 & 6/6/2008 the firm sent Urgent Device Notification letters to their customers. They notified their customers that a Field Service Engineer will inspect and repair the brake system.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution: USA including states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and WV and countries of Argentina, Australia, Austria, Bahamas, Belgium, Brazil, Canada, China, Colombia, Costa Rica, Czech Republic, Denmark, Egypt, England, France, Germany, Greece, China, India, Indonesia, Ireland, Iraq, Italy, Japan, Jordan, Korea, Lebanon, Libya, Lithuania, Malaysia, Malta, Mexico, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Saudi Arabia, South Africa, South Korea, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Venezuela.