Response 5.5/6.0 Spine System 7.0mm x 40mm Pedicle Screw Product Usage: Posterior non-cervical pedicle screw fixation
FDA Recall Z-0597-2018 · Posted None
Advertisement
Recalling FirmOrthoPediatrics Corp
Recall StatusTerminated
Product CodeOSH
Date Initiated2017-08-03
Date Terminated2019-05-08
Product Quantity24 units
Reason for Recall
The 7mm diameter uniaxial pedicle screws have been color anodized with the same color utilized on 5 mm diameter screws.
Firm Action
On August 4, 2017 an urgent medical device recall letter, titled "Response 5.5/6.0 Spine System - 7.0mm x 40mm UNIAXIAL PEDICLE SCREW" was issued to customers with instruction to quarantine affected product, which will be returned to the recalling firm for destruction. Questions or concerns can be directed to [email protected] or calling (574) 268-3692
Root Cause
Employee error
Distribution Pattern
US Nationwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.