Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
FDA Recall Z-0597-2026 · Posted 2025-12-09
Advertisement
Recalling FirmPhilips North America
Recall StatusOpen, Classified
Product CodeJAA
Date Initiated2025-10-27
Product Quantity110
Reason for Recall
Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm
stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
Firm Action
Philips will send a customer notification to the users of the affected systems to alert them of the issue
and provide the actions that should be taken to prevent risks to patients.
Philips will contact customers to schedule a time for a Philips Field Service Engineer (FSE) to visit their site and install the SSD spacer to resolve the issue.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
U.S.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.