Baxter Colleague CX Infusion Pump, Single Channel Volumetric Infusion Pumps. Product Code: 2M8161.

FDA Recall Z-0598-2012 · Posted 2012-01-11

Recall StatusTerminated
Product CodeFRN
Date Initiated2009-05-08
Date Terminated2012-01-03
Product Quantityone pump

Reason for Recall

There was no documentation to show the required tests were performed following the replacement of the battery and battery harness.

Firm Action

Baxter sent a service technician to the consignee location on 05/08/2009 to perform the battery check and discharge test. The pump was found to be within specification.

Root Cause

Process control

Distribution Pattern

Puerto Rico