Terminated

Merge PACS software. Product Usage: Merge PACS (Picture Archiving Communication System) is designed and marketed for soft copy reading, communication and storage of studies produced by digital modalities, including digital mammography.

FDA Recall Z-0599-2017 · Posted None

Recall StatusTerminated
Product CodeLLZ
Date Initiated2016-03-28
Date Terminated2021-04-23
Product Quantity534 sites potentially have the affected versions

Reason for Recall

The software did not show unviewed images when the last view was skipped. The physician may potentially misdiagnose when not all images are available for viewing.

Firm Action

The recalling firm issued a letter dated March 28, 2016 via e-mail or via certified mail to customers. The letter identified the affected product, problem and actions to be taken. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. The notification also informs the customer their response is required using the enclosed form and the return addressed envelope.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - US Nationwide including Canada.