Medtronic InSync Marquis, model 7277 Dual Chamber Implantable Cardioverter Defibrillator (ICD) with Cardiac Resynchronization Therapy devices are tiered therapy implantable arrhythmia control devices, which provide automatic detection of ventricular arrhythmias and delivery of user-selected therapies for the detected arrhythmia. Therapies are automatically delivered following detection or can be initiated via programming by the physician. These ICDs are multi-programmable and have the capability of transmitting via telemetry the programmed values, measured and collected data, and event markers.

FDA Recall Z-0600-05 · Posted 2005-03-11

Recall StatusTerminated
Product CodeNIK
Date Initiated2005-02-09
Date Terminated2006-12-28
Product Quantity7,704 total, (5,594 in US & 2,110 OUS)

Reason for Recall

Medtronic marquis family of ICD and CRT-D devices having batteries manufactured prior to December 2003 that may experience rapid battery depletion due to a specific internal battery short mechanism. Once a short occurs, depletion can take place within a few hours to a few days, after which there is a complete loss of device function.

Firm Action

Medtronic Sales Representatives contacted affected physicians starting 02/10/05 to deliver an Important Patient Management Information. The letter recommended patient managment options including quarterly follow-ups, programming Patient Alert to 'On-High', provide magnet to patients to check device status and inform patients that should they expereicnce warmth in the area surrounding the ICD to seek prompt care.

Root Cause

Other

Distribution Pattern

worldwide

Official Source

View FDA Recall →