Terumo Advanced Perfusion System 1, 4 inch diameter (small) Roller Pump; Model 801040.

FDA Recall Z-0600-2007 · Posted 2007-06-30

Recall StatusTerminated
Product CodeDTQ
Date Initiated2004-09-01
Date Terminated2009-07-07
Product Quantity1,484 of both models

Reason for Recall

The internal ribbon cable may loosen from the roller pump display board, resulting in loss of local control and/or local display.

Firm Action

The firm corrected units in the field or brought them back for evaluation and repair as complaints were received between 9/04 and 4/06. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that they will be visited and their units upgraded as replacement components become available.

Root Cause

Other

Distribution Pattern

Worldwide: USA, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Korea, Mexico, Malaysia, Singapore, South Africa, Taiwan, Thailand and United Arab Emirates.