Terminated

Synergy Pulse, Powered Flotation Therapy bed, Model number 5609. Designed to provide enhanced static and pulsation pressure relieving therapies. Provides sixteen pressure zones for customizing the support surface to each patient, and is capable of providing pulsation therapy.

FDA Recall Z-0600-2009 · Posted 2009-01-07

Recall StatusTerminated
Product CodeFNM
Date Initiated2008-10-09
Date Terminated2009-12-10
Product Quantity2,083 units (Total for all 4 units)

Reason for Recall

These devices have the potential to overheat.

Firm Action

All end users (those who either purchased the device and any customer that has rented) were notified by an Urgent Field Safety Notice on/about 10/31/2008. The letter informed the customers of the safety risk and list steps they can take to mitigate the risk until a Hill-Rom representative is able to modify the device. For the devices that were sold, the consignee of each device was notified by letter for the safety risk and what they can do to mitigate the risk until a Hill-Rom representative is able to modify the device. For the devices that are rented, an internal notification was sent via e-mail to Service Centers (Because the device is owned by the manufacturer, Hill-Rom, each Service Center in possession of the device will be the Consignee). Each Service Center will perform the correction on the units affected.

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution.