Terminated

Asante Conset Infusion Set with 110cm tubing & 6mm cannula packaged in 5-packs, singles,or inside a Combination Kit (FG-5246-4) Distributed by Asante Solutions, Sunnyvale, CA; Manufactured by Convatec; Lejre, Denmark.

FDA Recall Z-0600-2014 · Posted 2013-12-27

Recall StatusTerminated
Product CodeFPA
Date Initiated2013-11-19
Date Terminated2014-04-24
Product Quantity454 kits

Reason for Recall

Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the infusion set and providing patient with less insulin than intended.

Firm Action

Asante sent urgent medical device recall notifications letters to distributors and customers on November 19, 2013 by US mail, return receipt requested.

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution.