Asante Conset Infusion Set with 110cm tubing & 6mm cannula packaged in 5-packs, singles,or inside a Combination Kit (FG-5246-4) Distributed by Asante Solutions, Sunnyvale, CA; Manufactured by Convatec; Lejre, Denmark.
FDA Recall Z-0600-2014 · Posted 2013-12-27
Advertisement
Recalling FirmAsante Solutions, Inc.
Recall StatusTerminated
Product CodeFPA
Date Initiated2013-11-19
Date Terminated2014-04-24
Product Quantity454 kits
Reason for Recall
Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the infusion set and providing patient with less insulin than intended.
Firm Action
Asante sent urgent medical device recall notifications letters to distributors and customers on November 19, 2013 by US mail, return receipt requested.
Root Cause
Device Design
Distribution Pattern
Nationwide Distribution.