DPX-MD series, GE Lunar DEXA bone densitometers operating with enCORE software versions 12.0, 12.1, 12.2,13.0, & 13.1. The systems provide an estimate of BMD.
FDA Recall Z-0603-2010 · Posted 2010-01-27
Advertisement
Recalling FirmGe Healthcare
Recall StatusTerminated
Product CodeKGI
Date Initiated2009-02-01
Date Terminated2011-12-31
Product Quantity1,764 total
Reason for Recall
Failure to properly pre-indicate the technique factors to be used during a patient scan.
Firm Action
An Important Electronic Product Radiation Warning letter will be sent to customers. The letter will discuss the defect description and related hazards, the affected product details, safety instructions, and product correction. Customers are to contact their local GE Healthcare service representative with questions.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
Nationwide Distribution