Terumo Cardiovascular Custom Cardiovascular Procedure Kits containing an Overpressure Safety Valve, LN130B, in the Arterial Recirculation Line
FDA Recall Z-0603-2014 · Posted 2013-12-27
Advertisement
Recalling FirmTerumo Cardiovascular Systems Corporation
Recall StatusTerminated
Product CodeMNJ
Date Initiated2013-06-04
Date Terminated2014-10-01
Product Quantity24,227.
Reason for Recall
Over-Pressure Safety Valves not cleared with an indication for use on the arterial side of the extracorporeal bypass circuit.
Firm Action
Terumo Cardiovascular issued Customer notification letters on June 4, 2013. Accounts advised to: Review this notice and decide to continue to use or not use the Cardiovascular Procedure Kits as configured until a replacement kit with a one-way valve indicated for use on the arterial side of the perfusion circuit is available. If you have questions, contact Terumo CVS Customer Service: 800-521-2818.
Root Cause
Component design/selection
Distribution Pattern
Nationwide Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.