Spinal Concepts polyaxial screwdriver II, Manufactured by Spinal Concepts, Austin, TX, REF 781-2.

FDA Recall Z-0604-2011 · Posted 2010-12-10

Recalling FirmZimmer Inc.
Recall StatusTerminated
Product CodeKWQ
Date Initiated2005-12-01
Date Terminated2011-06-01
Product Quantity27

Reason for Recall

The devices may fracture, resulting in surgical delay or the need for intervention to remove fragments from the surgical site.

Firm Action

The affected instruments were removed and replaced in the U.S. Abbott Spine and Spinal Concepts instrument kits between December 2005 and October 2007. User accounts were not advised of the removal Zimmer has confirmed that none of these instruments are currently available for use in the U.S. For further information about this action, contact Dale Miller of Zimmer at 574 372-4962.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution, including US.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.