SOMATOM Definition AS, Computed tomography system Product Usage: SOMATOM Definition Flash systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.

FDA Recall Z-0604-2017 · Posted 2016-11-17

Recall StatusTerminated
Product CodeJAK
Date Initiated2016-10-19
Date Terminated2017-03-13
Product Quantity762 systems distrtibuted in U.S.

Reason for Recall

A problem can occur with the small light marker windows, used for the positioning laser and are integrated in the front cover of the SOMATOM Definition AS and Definition Flash systems, by either becoming loose or potentially dropping out. There is a risk that a person could touch rotating or electrical parts of the gantry if they were to reach through the window opening.

Firm Action

Siemens' sent a Customer Safety Advisory Notice letter dated October 19, 2016 to affected customers. The letter identified the affected product, problem, potential risks, how to avoid the potential risk related to this issue and actions to be taken.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide Distribution