Liko AB Uno Lifts, Powered (non-AC) Patient Lift: Uno 100 EM/EE: S/N 7090001 through 7096200 Uno 101: S/N 10001 through 11000 Production Dates: Manufactured prior to June 26, 2008 Uno 100 and 102 are battery powered mobile lifts, designed to be used in most of the common lifting situations, for example transfers between bed and wheelchair, to and from the toilet and bathtub or for lifting to and from the floor.

FDA Recall Z-0605-2009 · Posted 2009-01-27

Recall StatusTerminated
Product CodeFSA
Date Initiated2008-11-11
Date Terminated2016-02-04
Product Quantity24070 total units

Reason for Recall

Actuator of the lift has the potential to separate when the upper arm is fully extended.

Firm Action

Hill Rom notified customers with a Field Safety Notice, dated December 5, 2008 to notify customers of the issue, potential risks, and to inspect units per their instructions and remove units from service if the unit fails inspection. Hill-Rom will provide a permanent fix, using an "OuterTube" to be placed on all units.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution --- including USA and country of Canada.