Baxter INTERLINK System, Y-Type Solution Set, 78 (2.0 m), 2 Injection Sites, Male Luer Lock Adapter, 1C8581 s, Product Usage: C8581: For the administration of fluids from a container into the patients vascular system through a vascular access device. .
FDA Recall Z-0606-2016 · Posted None
Recalling FirmBaxter Healthcare Corp.
Recall StatusTerminated
Product CodeFPA
Date Initiated2015-11-30
Date Terminated2016-10-30
Product Quantity9600 units
Reason for Recall
Potential for dark loose particulate matter (400 microns or smaller) on the spike of the spike adapter or solution set.
Firm Action
Baxter sent an Urgent Product Recall letter by letter on 11/30/2015 to affected customers. The recall was extended to the consumer/user level. The letter instructed customers to locate and remove all affected product, contact Baxter Healthcare Center for Service to arrange for return of affected products and credit and complete the enclosed Baxter Customer Reply Form and return it to Baxter by either fax or scanned email.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide Distribution and the country of Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.