Discovery NM/CT 670 ES, model 5376204-70-57
FDA Recall Z-0607-2019 · Posted None
Advertisement
Recalling FirmGE Healthcare, LLC
Recall StatusCompleted
Product CodeKPS
Date Initiated2018-05-24
Product Quantity2 units
Reason for Recall
When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.
Firm Action
The firm initiated the recall by letter on 05/24/2018. The letter explained the issue and requested the UPS(s) be disconnected from the system until the system could be repaired.
Root Cause
Component change control
Distribution Pattern
CA, DC, OH, UT, WI, Canada