Ottobock Modular Polycentric EBS Knee Joint 3R60, 3R60=ST, 3R60=KD, 3R60=HD. Used for the prosthetic fitting of amputations of the lower. limb.

FDA Recall Z-0608-2016 · Posted None

Recall StatusTerminated
Product CodeISY
Date Initiated2016-01-04
Date Terminated2016-05-02
Product Quantity40 units

Reason for Recall

Otto Bock HealthCare GmbH is recalling Modular Polycentric EBS Knee Joint that were serviced between 27 April 2015 and 17 November 2015 due to a service error where the wrong Loctite was used during repair.

Firm Action

Starting Jan 4 2016, In coordination with the Manufacturer, the US Agent will notify customers via telephone call of the correction on behalf of the Manufacturer. The customers will be requested to return the affected devices back to the service center for correction. The devices will be returned to the customer following correction.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution.